STEEV - Stereotactic End-To-End Verification Phantom Patient - CIRS 038
For Comprehensive Testing of Stereotactic Radiosurgery Systems
PRODUCT NOT AVAILABLE
SRS commissioning and treatment verificationStereotactic Radiosurgery (SRS) necessitates a high degree of accuracy in target localization and dose delivery. Small errors can result in significant under treatment of portions of the tumor volume and overdose of nearby normal tissues. The CIRS Stereotactic End-to-End Verification Phantom "Steev" provides a means to check every step the patient will undergo in the treatment process-- from diagnostic imaging with CT, MR and PET, to treatment plan verification.
Accurate patient simulationSteev's anthropomorphic exterior allows for use of multiple positioning and fixation devises as used in clinical application. Internal details such as cortical and trabecular bone, brain, spinal cord, teeth, sinuses and trachea provide the most realistic clinical simulation to evaluate the challenging effects of complex intra-cranial and extra-cranial anatomies. Geometric and organic target inserts provide means for comprehensive image QA, geometric machine QA and TPS QA for increased confidence in system performance.
Dose measurements at isocenter and off isocenter positionsSteev accommodates a variety of interchangeable tissue equivalent inserts suitable for small field dosimetry including: micro-point and pin-point ion chambers, film, MOSFET, TLD, OSL (nanoDot™*) and 3D gel. When used in concert with the various imaging inserts, Steev provides the most comprehensive end-to-end testing and QA solution for SRS systems.
Features:- Perform End-to-End testing for commissioning as directed by AAPM TG-101
- Verify patient positioning using frame/frameless systems, head and shoulder masks or other positioning fixation devices
- Verify patient treatment plan in critical regions
- Perform geometric machine QA - Winston-Lutz isocenter verification tests and localization/repositioning with couch shift
- Perform IGRT QA procedure for X-ray and onboard kV and MV imagers including CBCT
- Assess image fusion, image transfer QA, accuracy verification and TPS testing with Multi-modality imaging capabilities (CT, MRI and PET)
- TPS Deformable Image registration algorithm accuracy QA


Axial T1 and T2 images of 50 cc Organic Target Insert aligned to external MR/CT fiducials

CT image of Spatial 3D Distortion Rectangular Insert
Specifications | |
Model 038 Includes |
|
Model 038 Optional Accessories | |
038-01 | Shoulders (additional 100 mm) |
038-02 | ABS vacuum formed cradle |
038-03-CvXX-XX |
Ion Chamber Dosimetry Kit:
|
038-04-CvXX-XX |
Variable Ion Chamber Position Dosimetry Kit:
|
038-05 | Film Cube for Single Film Dosimetry with Ø 30 mm target |
038-06 | TLD Dosimetry Cube |
038-08 | Multiple-OSL Dosimetry Target Insert |
038-09 |
Electron Density Cube with Real Water Electron Density plug (Water Equivalent Material Surrounding Removable Ø 1" Vial) |
038-10 | Winston Lutz Cube with Centroid & Offset Ø 5 mm targets |
038-11 | MRI/CT/PET Target Cube with Ø 30 mm target |
038-12 | MRI/CT/PET Target Cube with 25 cc Organic target |
038-13 | MRI/CT/PET Target Cube with 12.5 cc Organic target |
N/A | MR/CT/PET Spatial 3D Distortion Rectangular Insert |
Refer to separate CIRS cavity and plug code list for available chamber cavities. |