MRI-LINAC DYNAMIC PHANTOM - CIRS 008M
Image Acquisition - Treatment Planning - Dose Delivery

The CIRS MRI-LINAC Dynamic Phantom is designed for end-to-end commissioning, machine QA for hybrid MRI-LINAC Radiation Therapy systems, and verification of gated and adaptive treatment plans.

The Model 008M provides a means to assess MRI image quality, and to test tumor localization and motion-capturing devices. The phantom can be used to acquire moving or static target images for treatment planning and verification of gated or adaptive treatment delivery.

The phantom body, moving rod with target and extended rigid shaft connection are MRI-safe with 1.5T magnets. Motors and motion controller electronics are located at a safe distance from the magnet to ensure MRI image quality. The MRI-LINAC Dynamic Phantom utilizes CIRS Motion Control Software to provide precise, repeatable 3D target motion.

The phantom body approximates an average human body in size and shape. It is comprised of a plastic shell with a cylindrical thru hole to accommodate a moving rod with simulated target. The shell is filled with proprietary MRI signal-generating gel that allows for quick and easy phantom set-up without liquid motion artifacts. The phantom body is suspended on non-magnetic rails that allow for coil positioning and phantom registration to the couch.

Model 008M contains MRI QA targets that are compatible with ACR accreditation program recommendations for:

The moving insert with simulated target is a sealed, cylindrical rod filled with the same MRI signal-generating proprietary gel as the phantom body. The embedded MRI/CT target is made of MRI gel that provides contrast with respect to the background. This target can receive an ion chamber or other MRI-compatible dosimeters for treatment dose verification. There are laser marks machined on the outside of MRI shell body for both the MRI/CT target and MRI Distortion Grid. To facilitate MRI/CT Image fusion, there is a central fiducial aligned with the exterior laser marks, and in a known position with respect to 3D spatial distortion grid.

Features: User Friendly Motion Control

The CIRS MRI-LINAC Dynamic Phantom target motion is independently controlled using CIRS Motion Control Software. The graphical user interface provides an unlimited variety of motions while simplifying the operation of Model 008M to an intuitive level. Patient specific motion profiles are easily imported and there is no need to make hardware adjustments or have special programming skills.


Specifications
Motion Controller
Overall Dimensions 67 cm x 32 cm x 20 cm
Overall Weight 10kg (22lb)
Amplitude, IS ± 25mm
Amplitude, AP/LR ± 10mm
Motion Accuracy ± 0.1mm
Cycle Time 1 - ∞ (adjusted based on amplitude)
Waveforms sin (t), 1-2cos4(t), 1-2cos6(t), sawtooth, sharkfin
Phantom Body
Overall Dimensions 25.6cm x 32cm x 20cm
Overall Weight 17kg (38lb)
Moving Rod
Overall Dimensions 32.7cm x Ø 6.3m
Overall Weight 1.5kg


MRI-LINAC DYNAMIC PHANTOM - CIRS 008M

CIRS 008M - 3D Grid MRI Distortioned vs CAD
3D Grid MRI Distortioned vs CAD

CIRS 008M - Target MRI Distortioned vs CAD Model
Target MRI Distortioned vs CAD Model

CIRS 008M - MRI Main Views
MRI Main Views
CIRS 008M - Top, Perspective, Front and Side Views
Top, Perspective, Front and Side Views
CIRS 008M - Central Fiducial Detail
Central Fiducial Detail
CIRS 008M - Motion of Moving Rod Insert
Motion of Moving Rod Insert
Copyright ©
Warning: date(): It is not safe to rely on the system's timezone settings. You are *required* to use the date.timezone setting or the date_default_timezone_set() function. In case you used any of those methods and you are still getting this warning, you most likely misspelled the timezone identifier. We selected the timezone 'UTC' for now, but please set date.timezone to select your timezone. in /home/supertechxray/public_html/phpincludes/copyright.php on line 2