Multipurpose & Endoscopic Ultrasound Phantom ATS 570
The Model 570 Multipurpose & Endoscopic phantom is an easy, comprehensive means of evaluating imaging systems with an operating frequency range of 2.25 to 7.5 MHz.
The phantom is designed with a combination of monofilament line targets for distance measurements and tissue mimicking target structures of varying sizes and contrasts.
Due to the acoustic similarity of the background material and the target structures, artifacts caused by distortion, shadowing and enhancement have been eliminated. Four
gray scale targets ranging in contrast from +6 to -3 dB are provided to evaluate the system’s displayed dynamic range and gray scale processing performance.
The Model 570 offers a new and improved scan surface design to easily accommodate linear, sector, endoscopic probes and mechanical sector probes such as used for rectal scanning.
Scan Surface
|
|
|
|
|
Standard Scan Surface |
Endoscopic Scan Surface |
Scanning Well |
|
Tests Performed:
- Dead Zone or Ring-Down (A1 & A2)
- Horizontal Measurement Calibration Linear (C1 & C2)
- Horizontal Measurement Calibration Sector (C1 & C2)
- Sensitivity/Penetration (F)
- Vertical Measurement Calibration (B1 & B2)
- Image Uniformity
- Functional Resolution (F1, F2)
- Focal Zone (B1, B2, or F)
- Axial & Lateral Resolution (E1, E2, E3)
- Contrast, Gray Scale (G1, G2, G3)
|
Uses:
- Quality Assurance
- Routine Clinical Performance
- In-process Manufacturing
- Final Q.A. Inspection
- Research & Development
- Teaching
|
Specifications
|
General
|
|
Overall Dimensions |
27 x 21.5 x 9.6 cm |
|
Housing Material |
PVC |
|
Scan Surfaces |
2 |
|
Scan Wells |
1 |
|
Weight |
4.55 Kg (10 lb) |
|
|
Tissue Mimicking Material
|
|
Type |
Urethane Rubber |
|
Freezing Point |
< -40°C |
|
Melting Point |
> 100°C |
|
Attenuation Coefficient |
0.5 dB/cm/MHz ± 5.0% |
|
Speed of Sound |
1450 m/s ± 1.0% at 23° |
|
|
Line Targets
|
Material
Diameter
Position Tolerance |
Monofilament Nylon
.12 mm
± 0.1 mm
|
Dead Zone Groups
Lateral Displacement
Interval Spacing
Scan Surface Depth
|
Standard
Endoscopic
|
|
|
|
|
Vertical Groups
|
Number of Targets
Interval Spacing
Depth
|
|
|
|
Horizontal Linear
Groups
|
Number of Targets
Interval Spacing |
|
Scan Surface Depth |
Standard
Endoscopic |
|
|
|
|
Axial-Lateral Resolution Groups
|
|
|
|
|
Number of Targets
|
|
|
Interval Spacing
|
|
|
|
Scan Surface Depth
|
|
|
|
|
Horizontal Linear
Groups
|
Number of Targets
Interval Spacing
|
|
Scan Surface Depth
|
Standard
Endoscopic
|
|
|
|
|
6 |
|
5, 4, 3, 2, 1 mm |
|
40 mm |
60 mm |
40 mm |
|
Anechoic Target Structures |
F1 |
F2 |
Gray Scale Target Structures |
G1
Standard |
G2
Endoscopic |
G3
Scan Well |
|
Type |
Non-echogenic, cylindrical |
Type |
Echogenic, Cylindrical |
|
Size |
8, 4, 2 mm |
6 mm |
Diameters |
10 mm |
10 mm |
6 mm |
|
Number of Targets |
27 |
1 |
Scan Surface Depth |
30 mm |
20 mm |
30 mm |
|
Depth |
10 - 170 mm |
30 mm |
Number of Targets |
4 |
4 |
2 |
|
Interval Spacing |
10 & 20 mm |
30 mm |
Contrast relative to background material (dB) |
+6, +3, -3, -6 |
+3, -3 |