Resolution – Penetration Ultrasound Phantom ATS 534
The Model 534 tissue-mimicking™ phantom is designed to evaluate the performance of diagnostic ultrasound imaging systems. These phantoms
contain 81 anechoic, cylindrical target structures of varying sizes. The targets are positioned in-line vertically, to permit an entire target group to be displayed
in one view. Due to the acoustic similarity of the background material and the anechoic targets, artifacts caused by distortion, shadowing or enhancement have been eliminated.
The Model 534 utilizes a rubber-based™ material with a Lifetime warranty.
Uses
- Quality Assurance
- Basic Clinical Performance
- In-process Manufacturing
- Teaching
|
Tests Performed
- Sensitivity
- Focal Zone
- Functional Resolution
- Image Uniformity
|
|
SPECIFICATIONS: Model 534
|
|
General
|
|
Overall Dimensions: |
20.3 x 23.3 x 9.5 cm |
|
Scan Surfaces: |
4 |
|
Scan Surface Dimensions: |
14.5 x 7.5 cm |
|
Scan Surface Dimensions: |
17.5 x 7.5 cm |
|
Weight: |
3.1 Kg |
|
Tissue Mimicking Material
|
|
Type: |
Urethane Rubber |
|
Freezing Point: |
<-40°C |
|
Melting Point: |
>100°C |
|
Speed of Sound: |
1450 m/s ± 1.0% @ 23°C |
|
Attenuation Coefficient: |
0.5 dB/cm/MHz ± 5.0% |
|
Anechoic Target Structures
|
|
Type: |
Non-echogenic, cylindrical |
|
Diameters: |
2, 3, 4, 6, 8, 10 mm |
|
Depth: |
1 - 18 CM |
|
Includes Carrying Case & Built-in Scan Wells. |
|
|