Pro-NM NEMA NU 2 Resolution
PET Spatial Resolution Phantom - Diagnomatic by Pro-Project - 08-109

Pro-NM NEMA NU 2 Resolution from Diagnomatic by Pro-Project supports measurements of reconstructed point-spread functions from compact radioactive sources.

A chair-shaped object made of two parallel transparent flat elements connected with a few rods and nested on a rectangular transparent base.
A chair-shaped object made of two parallel transparent flat elements connected with a few rods and nested on a rectangular transparent base.

Click a photo or illustration to enlarge it.

Models and Ordering

Part numberModel / configuration
08-109Pro-NM NEMA NU 2 Resolution

Specifications

  • five capillaries with an inside diameter of 1 mm and an outside diameter of less than 2 mm
  • in the transverse direction the capillary is positioned at 1 cm (to represent the center of the FOV, but positioned to avoid any possible inconsistent results at the very center of the FOV), 10 cm, and 20 cm from the center of the plane
  • middle of the capillaries is marked to provide guidance while the radionuclide is introduced
  • markers and convenient base with a spirit level for easy and accurate positioning
  • optional carrying case (08-109-CS)

Features and Test References

  • complies with:
    - International Standard: Radionuclide imaging devices – Characteristics and test conditions – Part 1: Positron emission tomographs, International Electrotechnical Commission (IEC), 61675-1, Geneva, Switzerland, 1998.
    - Performance Measurements of Scintillation Cameras, NEMA Standards Publication No. NU2, National Electrical Manufacturers Association (NEMA), Washington, D.C., 2001.
    - NEMA 2007/IEC2008
    - NEMA 2012/IEC 2008
    - NEMA Standard NU 2-2018
  • CE marked
  • Includes a manual covering the tests, result evaluation and recording.