Pro-NM NEMA NU1 RSR
SPECT Resolution and Alignment Phantom - Diagnomatic by Pro-Project - 08-116
Pro-NM NEMA NU1 RSR from Diagnomatic by Pro-Project holds three capillary line sources for reconstructed resolution measurements without scatter and system alignment checks.
Three thin-walled glass capillaries, each with an internal diameter of approximately 1 mm, are positioned on a dedicated support plate.
Click a photo or illustration to enlarge it.
Models and Ordering
| Part number | Model / configuration |
|---|---|
| 08-116 | Pro-NM NEMA NU1 RSR |
Specifications
- Aluminum Plate
- Length: 400 mm
- Width: 200 mm
- Height: 6 mm
- Capillaries
- Capillary length: 200 mm
- Capillary diameter: inner 0.87 mm (outer 1.5 mm)
- Capillaries plugging clay
- Phantom weight: 1300 g
Accessories
- Additional set of capillaries and clay plug
- Optional carrying case
Features and Test References
- Complies with:
- NEMA Standards Publication (NU 1-2023) Performance Measurements of Gamma Cameras - CE marked
- Includes a manual covering the tests, result evaluation and recording.
Additional Product Illustrations