Pro-MAM CESM
Contrast-Enhanced Mammography Phantom - Diagnomatic by Pro-Project - 03-701

Pro-MAM CESM from Diagnomatic by Pro-Project uses iodine targets and breast-equivalent backgrounds to assess contrast-enhanced spectral mammography images.

The 10 mm target plate represents a 50/50 glandular/fat mixture. Two target groups each contain iodine concentrations of 0.25, 0.5, 1.0 and 2.0 mg/cm³ plus a glandular-tissue lesion target. A lower module provides separate backgrounds representing 100% glandular tissue and 100% adipose tissue.

A white object consisting of a wall with a semicircular base, two layers and a lid.
A white object consisting of a wall with a semicircular base, two layers and a lid.
A white object consisting of a wall with a semicircular base and two layers, the cover has been removed. The top pink layer contains two groups of five elements.
A white object consisting of a wall with a semicircular base and two layers, the cover has been removed. The top pink layer contains two groups of five elements.
A white object consisting of a wall with a semicircular base and a layer, the cover and first layer have been removed. The layer is split in half, representing two different tissues.
A white object consisting of a wall with a semicircular base and a layer, the cover and first layer have been removed. The layer is split in half, representing two different tissues.

Click a photo or illustration to enlarge it.

Models and Ordering

Part numberModel / configuration
03-701Pro-MAM CESM

Product highlights

  • CESM routine quality control
  • Assessment of the quality of subtracted images
  • Contains iodine concentrations in a clinically relevant range
  • Represents types of breasts with the majority of glandular tissue or fat tissue
  • Object visibility assessment is similar to clinical conditions

Specifications

  • phantom thickness: 55 mm
  • the outer layers (top and bottom) are the equivalent of 10 mm thick adipose tissue
  • the middle layer, 25 mm thick, consists of half of a material that simulates adipose tissue and half of a material that simulates glandular tissue
  • a module with a thickness of 10 mm contains 2 groups of test objects

Features and Test References

  • complies with:
    - Klausz, R, et al. Introduction of a Comprehensive Phantom for the Quality Control of Contrast Enhanced Spectral Mammography. 2018. Poster No.: ECR 2018, DOI: 10.1594/ ecr2018/C-2647
  • CE marked
  • Includes a manual covering the tests, result evaluation and recording.