Pro-MAM CBCT - 03-702
Breast CBCT Quality Assurance Phantom - Diagnomatic by Pro-Project

Pro-MAM CBCT is a modular phantom for acceptance testing and routine quality control of breast cone-beam CT systems. Its separate sections support evaluation of microcalcifications, contrast and image uniformity in a tissue-equivalent background.

Assembled Pro-MAM CBCT 03-702 breast phantom in pink tissue-equivalent material Pro-MAM CBCT - Model 03-702

Click any photo to enlarge it.

Technical Specifications

Model03-702
Diameter130 ± 0.5 mm
Height115 ± 2 mm
ShapeSemi-ellipsoid
MaterialBR 20/80 tissue-equivalent mixture: 20% glandular / 80% fat
Test sectionsCalcification clusters, contrast resolution and uniformity
Product markingCE marked

Calcification Clusters Module

Microcalcification groups are embedded in a glandular/fat background. Calcification point diameters are 320, 290, 270, 240, 220 and 160 µm.

Pro-MAM CBCT calcification clusters module with small test points in the tissue-equivalent background Calcification clusters module

Contrast Resolution Module

High- and low-contrast targets support contrast-to-noise ratio (CNR) and low-contrast detectability measurements. A water target is also provided.

Pro-MAM CBCT contrast resolution module with targets of different sizes Contrast resolution module

Uniformity Module

The semi-ellipsoidal section is used to assess image uniformity and artifacts.

Pro-MAM CBCT uniformity module showing its smooth tissue-equivalent test section Uniformity module

Accessories and Documentation

Carrying handles and a transport case are optional. The supplied manual explains test procedures, evaluation and recording of results. Technical specifications can be adapted to customer requirements.

Download the Pro-MAM CBCT product card (PDF)

Copyright © 2026 Supertech, Inc. All rights reserved.