Pro-MAM CBCT - 03-702
Breast CBCT Quality Assurance Phantom - Diagnomatic by Pro-Project
Pro-MAM CBCT is a modular phantom for acceptance testing and routine quality control of breast cone-beam CT systems. Its separate sections support evaluation of microcalcifications, contrast and image uniformity in a tissue-equivalent background.
Click any photo to enlarge it.
Technical Specifications
| Model | 03-702 |
|---|---|
| Diameter | 130 ± 0.5 mm |
| Height | 115 ± 2 mm |
| Shape | Semi-ellipsoid |
| Material | BR 20/80 tissue-equivalent mixture: 20% glandular / 80% fat |
| Test sections | Calcification clusters, contrast resolution and uniformity |
| Product marking | CE marked |
Calcification Clusters Module
Microcalcification groups are embedded in a glandular/fat background. Calcification point diameters are 320, 290, 270, 240, 220 and 160 µm.
Contrast Resolution Module
High- and low-contrast targets support contrast-to-noise ratio (CNR) and low-contrast detectability measurements. A water target is also provided.
Uniformity Module
The semi-ellipsoidal section is used to assess image uniformity and artifacts.
Accessories and Documentation
Carrying handles and a transport case are optional. The supplied manual explains test procedures, evaluation and recording of results. Technical specifications can be adapted to customer requirements.