Supertech ®
Thousands of Products
Serving You is Our Pleasure
Product Categories
Small Parts Phantom ATS 551
Model 551 provides a comprehensive means of evaluating the performance of a diagnostic ultrasound system at frequencies ranging from 7.5 to 15 MHz. It is designed with a combination of monofilament line targets and tissue mimicking cylindrical targets of varying sizes and contrasts. The monofilament line targets have a diameter of 0.05 mm, to optimize the displayed image at frequencies of 7.5 MHz or greater.
Four groups of line targets are provided to evaluate the vertical and horizontal calibration measurements, dead zone and axial-lateral resolution. The Model 551 is constructed of our rubber-base tissue mimicking material to provide a phantom which is accurate, durable and maintenance free.
**Lifetime warranty**
Tests Performed
• Vertical Measurement
Calibration
• Horizontal Measurement
Calibration
• Axial Resolution
• Lateral Resolution
• Sensitivity
• Focal Zone
• Functional Resolution
• Image Uniformity
• Gray Scale
• Dead ZoneUses:
• Quality Assurance
• Routine Clinical Performance
• In-Process Manufacturing
• Final Q.A. Inspection
• Research & Development
• Teaching
Specifications
General Overall Dimensions 28.0 x 11 x 11cm *
Weight 3.1 Kg*
Housing Material Acrylic
Wall Thickness 1.0 cm*
Scan Surfaces 1
Scan Surface Dimensions 25 x 8 cm
Tissue Mimicking Material Type Urethane rubber
Freezing Point < -40°C
Melting Point > 100°C
Attenuation Coefficient 0.5 dB/cm/MHz ± 5.0%**
Speed of Sound 1450 m/s ±1.0% at 23°
Line Targets Material Monofilament Nylon Diameter 0.05mm ± 5% Position Tolerance ± 0.1 mm Vertical Number of Targets 10 Interval Spacing .5 cm Depth .5 to 5.0 cm Horizontal Number of Targets 9 Number of Groups 2 Interval Spacing .5 cm Depths Linear Array
Sector Array2.0 & 4.0 cm
2.5 & 4.5 cmDead Zone Lateral Displacement 5.0 mm Interval Spacing 1.0 mm Depth 1.0 - 4.0 mm Axial-Lateral Resolution Number of Groups 2 *Nominal dimensions
**Other attenuations are available upon request
Includes Carrying Case & Built-in Scan wells.
Supertech ®
Address:
P.O. Box 186
Elkhart, IN 46515 USA
Sales: 800-654-1054
Phone: 574-264-4310
Fax: 574-264-9551
Email: sales@supertechx-ray.com