Tissue Equivalent Calibration Standard
Quantitative Ultrasound Phantom CIRS 063
The Model 063 QUS Phantom provides a linear response of Broadband Ultrasonic Attenuation (BUA) in the diagnostic frequency range for assessment of bone quality.
Features
Linear response in the diagnostic frequency range
Can be molded into any shape (custom manufacturing)
Mimics calcaneus bone
Proven construction methodology
Known material properties permit phantom to be used as a calibration tool with various QUS systems
Model # Model
06301
Quantitative Ultrasound Phantom Normal Heel
06302 Quantitative Ultrasound Phantom Osteoporotic Heel
06303
Wedge Adapter for Quantitative Ultrasound Phantom
(For Hologic Systems)Model 063 Specifications
Material: Proprietary Urethane
Thickness: 36 mm
Speed of Sound Range: 1500-1600 m/s at room temperature
Test Conditions in CIRS Accoustic Laboratory:
Water bath
Transducer center frequency = 0.5 MHz
Sampling frequency = 20 million samples/sec.
Frequency range for BUA evaluation = 0.25 - 0.55 MHz
Temperature = 21.3 °C