General Purpose Multi-Tissue Ultrasound Phantom
Model 040
The standard for ultrasound quality assurance.
Two Phantoms in one case.
Model 040
The CIRS series of ultrasound phantoms, unlike human subjects or random scannable materials, offers a reliable medium which contains specific, known test objects for repeatable qualitative assessment of ultrasound scanner performance over time.
This phantom is constructed from the patented solid elastic material, Zerdine®(1). Zerdine®, unlike other phantom materials on the market, is not affected by changes in temperature. It can be subjected to boiling or freezing conditions without sustaining significant damage. Zerdine® is also more elastic than other materials and allows more pressure to be applied to the scanning surface without subsequent damage to the material. At normal or room temperatures the Zerdine® material found in the Model 040 will accurately simulate the ultrasound characteristics found in human liver tissue.
The Model 040 was designed to allow for assessment of uniformity, axial and lateral resolution, depth calibration, dead zone measurement, and registration within two different backgrounds of 0.5 and 0.7 dB/cm/MHz.
Specifications: Material: Zerdine®(1),
solid elastic water-based polymer
Freezing Point: 0 C
Melting Point: Above 100 CAttenuation Coefficient:
0.5 dB/cm/MHz
0.7 dB/cm/MHzSpeed of Sound: 1540 m/s
Scanning Well: 1 cm deep
Scanning Membrane: Saran
Targets:
Material: Nylon
Monofilament
Wire Diameter: 0.1 mmVertical Plane Targets
Number of Groups: 1
Number of Targets Per Group: 16
Depth Range: 18 cm
Spacing: 1 cmHorizontal Plane Targets:
Number of Groups: 2
Number of Targets: 4 and 7
Depth Range: 3 cm and 9 cm
Spacing: 1 cm and 2 cmResolution Targets:
Number of Arrays: 4
Depths: 2.5 cm, 6 cm and 10 cm
Axial Intervals: 0.5, 1 , 2, 3, 4, and 5 mm
Horizontal Intervals: 1, 2, 3, 4, and 5 mmCystic Targets:
Number of Targets: 4
Diameter of Targets: 2, 4, 6 and 8 mm
Depth of Targets: 2, 4, 6, and 8 cm
Attenuation: <0.07dB/cm/MHz
Speed: 1540 m/s
Contrast: anechoicHigh Contrast Targets:
Number of Targets: 4
Diameter of Targets: 2, 4, 6 and 8 mm
Depth of Targets 2, 4, 6, and 8 cm
Attenuation: 1.0 dB/cm/MHz
Speed: 1540 m/s
Contrast: +15 dB v.s. backgroundHigh Density Target:
Material: PMMA
Diameter: 1/16"
Depth: 6 cm
Phantom comes with detachable scanning wells to accommodate large sector probes and small endocavity probes. It is packaged in a hermetically sealed, air tight, rugged carrying case.