Supertech ®
Thousands of Products
Serving You is Our Pleasure
Product Categories
Tissue-Equivalent Phantom for Mammography
CIRS 011A
(Other Models listed at bottom of this page)A Refined Quality Assurance Tool for Today's Advanced Imaging Systems
Proven simulation technology enables the use of tissue-equivalent, realistically-shaped phantoms for mammographic quality assurance.
The resin material mimics the photon attenuation coefficients of a range of breast tissues. Average elemental composition of the human breast being mimicked is based on the individual elemental composition of adipose and glandular tissue reported by Hammerstein.
Attenuation coefficients are calculated by using the "mixture rule" and the Photon Mass Attenuation and Energy Absorption Coefficient Table of J.H. Hubbell. The CIRS Model 011A Breast Phantom contains targets that are engineered to test the threshold of the new generation of mammography machines. Model 011A is 4.5 cm thick and simulates an average glandular tissue composition.
The Model 011A was designed to test the performance of any mammographic system. Objects within the phantom simulate calcifications, fibrous calcifications in ducts and tumor masses. Test objects within the phantom range in size from those that should be visible on any system to objects that will be difficult to resolve on the best mammographic systems.
CIRS mammography phantoms are also manufactured in 4 cm, 5 cm and 6 cm thicknesses with various glandular equivalencies.
The methodology and design of these phantoms was developed by Dr. Panos Fatouros and his associates at the Medical College of Virginia.
Specifications:
Line pair target
1. 20 lp/mm
Ca CO3 specs
grain size (mm)
2. .130
3. .165
4. .196
5. .230
6. .275
7. .400
8. .230
9. .196
10. .165
11. .230
12. .196
13. .165
Step Wedge
1 cm thick
14. 100% gland
15. 70% gland
16. 50% gland
17. 30% gland
18. 100% adipose
Nylon Fibers
diameter size (mm)
19. 1.25
20. 0.83
21. 0.71
22. 0.53
23. 0.30
Hemispheric Masses
75% glandular/ 25% adipose, thickness (mm)
24. 4.76
25. 3.16
26. 2.38
27. 1.98
28. 1.59
29. 1.19
30. 0.90
Optical Density
31. reference zone
Edge of Beam
32. localization target
Phantom Body
Length 12.5 cm
Width 18.5 cm
Height 4.5 cm
Material Epoxy
Also Included
30x handheld microscope
Mammography QA docu- ments
for recording image evaluations
and scores
Technical manual
Carrying case sold separately.
References:
1. Skubic S.E., Fatouros PP. Abosorbed Breast Dose: Dependence on Radiographic
Modality and Technique, and Breast Thickness. RADIOLOGY,1986, 161:263-270.
2. Fatouros PP, Skubic S.E., Goodman H. The Development and Use of Realistically Shaped, Tissue- equivalent Phantoms for Assessing the Mammographic Process.
RADIOLOGY, 1985 157(p):32.
Available Models:
Model # Description 010A Tissue Equivalent Mammography Phantom, 5.0cm Thick 30%
Gland 70% Adipose010B Tissue Equivalent Mammography Phantom, 4.0cm Thick 50%
Gland 50% Adipose010C Tissue Equivalent Mammography Phantom, 6.0cm Thick 20%
Gland 80% Adipose010D Tissue Equivalent Mammography Phantom, 5.0cm Thick 50%
Gland 50% Adipose011A Tissue Equivalent Mammography Phantom, 4.5cm Thick 50%
Gland 50% Adipose
Supertech ®
Address:
P.O. Box 186
Elkhart, IN 46515 USA
Sales: 800-654-1054
Phone: 574-264-4310
Fax: 574-264-9551
Email: sales@supertechx-ray.com