ECC Digital X-ray Pulse Counter 8700
Direct measurement of Wave Pulse & Exposure Time

The Model 8700 X-Ray Pulse Counter/ Exposure Time Meter is used to measure the time or duration of radiation output produced by a wide variety of x-ray generators. The x-ray sensor in the 8700 allows direct measurement of exposure from the x-ray head. Pulses produced by half-wave and full-wave x-ray are measured as 60 or 120 pulses per second. For DC, capacitor discharge and 3-phase x-ray, the 8700 measures the exposure time in milliseconds.

When testing x-ray timers and controls, the time of relay contact can be measured using the AC input feature.

The Model 8700 is "self-resetting". There is no need to reset the instrument after each reading. The reading is stored after each exposure, holding until the next exposure.

New and Improved!

We have recently made significant changes to the design of our Model 8700 to improve the function and accuracy. No changes were made regarding the actual operation, so the units work the same as existing models.

However, the following performance improvements were made:

  • Wider Range

  • Increased sensitivity allows use with lower power x-rays

  • Wider dynamic range allows use with higher power x-rays

  • More accuracy on DC/Multiphase high power x-rays

  • Measurement ends at 70% of peak value eliminating inaccuracy due to slow tail.

  • New circuitry saves production time and parts cost, allowing us to keep our competitive prices.

  • Low power circuitry improves battery life.

 

Features

  • Small hand-held size

  • AC or DC X-Rays

  • New high accuracy circuit

  • Self-resetting

  • Battery operated

  • Two year warranty

  • Direct exposure reading or AC input

  • Use for Medical or Dental X-Rays

  • Remote Sensor option

  • No periodic calibration needed

  • Custom carrying case option

  • Durable ABS plastic case

  • Typical Applications

  • Measurement of Exposure

  • Quality Assurance

  • Verification of X-ray Exposure Time

  • Calibration of X-ray Timer

  • Direct Measurement of Timer Accuracy

  • Troubleshooting and Repair of x-ray

 

Specifications

ACCURACY:     AC - one count
                        DC- 2% +/-1 millisecond

SENSITIVITY:    X-ray input: 40kVp, 5mA at 5 cm from top surface of
                        case,  pointed to target
                        AC Input: 90 VAC minimum

RANGE:     9999 Pulses
                 9999 Milliseconds

DISPLAY:     0.4" Liquid Crystal

POWER:     Internal 9 Volt Battery, Low Battery Indicator

BATTERY LIFE:   48 hours continuous, typically 9 months of normal use

CONTROLS / INDICATORS

Three position switch:     Pulse, Off, Milliseconds

4 digit liquid crystal display:     0.4" characters, Low battery indicator
                                                 Oscilloscope output
                                                 Remote Sensor Input

PHYSICAL SIZE:     80 x147x 40mm, 3.15 x 5.8 x1.6 inches

CONNECTIONS:   None required for direct exposure measurement
                            2 tip jacks for AC input
                            Jack for Remote Sensor Option
                            Jack for Oscilloscope Output

 

X-ray Pulse Counter / Exposure Time Meter 8700

OPTIONAL ACCESSORIES

Remote Sensor with 3 meter cable allows operation of Pulse Counter away from x-ray 8700RS
Padded Cordura Carrying Case Carrying case specifically designed for instrument 8700CC
Oscilloscope Interface Cable connects ‘Scope’ output to standard BNC oscilloscope input. Allows radiation output waveform to be viewed on an oscilloscope 8700SC
Hard shell Carrying Case 8700HC

 

 

    