AtomlabTM 400 Dose Calibrator
Designed for facilities receiving unit doses including PET and Beta
- Pre-programmed for 88 most commonly used radionuclides
- Large, easy-to-read backlit LCD
- Small footprint economizes workspace
- Ultra-fast response
- Automatic range selection; ranges up to 40 curies of Tc-99m or 10 curies of F-18
- Displays in curies or becquerels
- Remote Ionization Chamber
- Remote Ionization Chamber
- Self-Diagnostic Software
- Desktop or wall mount display
- Two-year warranty
- RS-232 bi-directional serial communications port
Note: New low pressure chamber. Send air or ground as standard goods.
The Atomlab™ 400 provides fast, accurate radionuclide activity measurements with performance that easily complies with the most stringent regulatory requirements.
The unit is simple to operate. There is a routine list of ten pre-programmed isotopes plus another seven that can be user defined. Any four isotopes will be displayed at a time. In addition, there is a library containing 88 isotopes listed alphabetically, including Y-90 and Sr-89.
Activity is displayed on a LCD panel in either curie or becquerel units. Background correction is performed at the touch of a button. Range selection is automatic.Activity measurements are performed by a microprocessor-controlled electrometer located within the detector assembly of the ionization chamber. The chamber is shielded with .25" (6.3 mm) lead. It can be located up to eight feet away from the display unit. Chamber bias is generated within the display unit by an electronic high voltage supply, eliminating the need for expensive battery changes.
The RS-232 port enables the Atomlab 400 Dose Calibrator to communicate with most commercially available nuclear medicine management systems.
Specifications
Isotope Selection Keys: Ten pre-programmed – Tc-99m, Co-57, Cs-137,
I-131, In-111, Ga-67, Xe-133, I-123, Tl-201, and Mo-99; seven additional keys for user-set isotopes; two new isotope keys and a full alphabetical list of 88 isotopes.Activity Range: 0.1 µCi to 40Ci (.001 mBq to 1500 GBq) of Tc-99m
Energy Range: 25 keV to 3 MeV photons
Response Time: One to two seconds for doses greater than 200 uCi; three
seconds for doses greater than 20 uCi; 50-100 seconds below 20 uCi of
Tc-99m with default threshold, threshold adjustable to reduce counting timeDetector Linearity: ±1% or 0.2 µCi, whichever is greater
Electrometer Linearity: ±1% or 0.2 µCi, whichever is greater
Electrometer Accuracy: ±1% or 0.2 µCi, whichever is greater
Overall Accuracy: ±3% or 0.3 µCi, whichever is greater; overall accuracy is affected by such factors as the accuracy of the specific source calibration, geometric variations due to sample volume or configuration, detector linearity, electrometer accuracy and readout accuracy
Repeatability: ±0.3% above 1 mCi short term (24 hr); 1% long term (one yr); exclusive of background
Digital Calibration Dial: Four-digit LED dial display with increment/decrement keys to change the value; range is from 0.0 to 999.9
Detector: Well-type pressurized ionization chamber, with Argon fill gas; well opening 2.75" (7 cm), well depth 10.25" (26 cm)
Chamber Gas Pressure: 149KPa gauge (21.6 psig) at 20 degrees C or 250KPa absolute (36.3 psia) at 20 degrees C. IATA regulation 3.2.2.4 Exempts Gases of Division 2.2 from Dangerous Goods Regulations when transported at pressure less than 200KPa gauge (29 psig) at 20 degrees C. Device is shipped standard goods.
Detector Shielding: .25" (6.3 mm) lead on all sides except top well opening; supplementary shielding available
Chamber Bias: 355 ± 5 volts
Environmental Operating Conditions: Temperature: 0-40° C, Humidity: 0-90% rh, non-condensingPower Requirements: 100 to 240 VAC, 0.6 – 0.3 amps, auto switching; APS Power Supply (APS22ES-150160), for medical use.
Line Frequency: 50/60 Hz; detachable line cord; built-in EMI filter and transient suppression
Detector and Interface Cables: 8' (243 cm) long, six conductor cables (two carry power, two for chassis ground, two carry serial data for digital I/O)
Display Unit:
Dimensions: 6.75" w x 6" depth x 5" h (17.1 x 15.3 x 12.7 cm)
Weight: 3.6 lb (1.64 kg); desktop or wall mountable
Detector Unit:
Dimensions: 6" dia x 15.5" h (15.24 x 39.37 cm)
Well I.D.: 2.75" dia x 10.5" h (7 x 26.7 cm)
Well I.D. with Liner: 2.5" dia x 10.25" h (6.35 x 26 cm)
Lead Shielding: .25" thick (6.3 mm)
Weight: 35 lb (16 kg)
Approvals: ETL to UL 60601-1 and cETL to CAN/CSA C22.2 No. 601-1-M90
Warranty: Two Year
086-335 Dose Calibrator Atomlab™ 400, 100-240 VAC Includes: smart display, ionization chamber, rs-232 port, Vial/syringe dipper and well insert.
Related: 086-336 Chamber, Dose Calibrator 086-338 Dose Caibrator Shielding Rings, Interlocking 2" Lead. For additional protection from High Energy activity.
086-423 Moly Shield, Vial, .3" Lead 086-435 Moly Shield, Syringe, .3" Lead 086-509 Lineator 086-334 Cable, European to wall outlet Replacement: 086-241 Well Insert 086-242 Vial/Syringe Dipper
Supertech ®
Address:
P.O. Box 186
Elkhart, IN 46515 USA
Sales: 800-654-1054
Phone: 574-264-4310
Fax: 574-264-9551
Email: sales@supertechx-ray.com
Support: support@supertechx-ray.com