8000 NERO® mAx Victoreen® X-Ray Test Device
Non-invasive evaluator of radiation outputs
the most difficult machines 100 kHz sampling speed captures data from
0.5 kV or 1% accuracy from 22 to 160 kV
Measures kVp average, kV effective, kV peak, time, exposure or rate, mA or mAs, HVL, exposure/frame, and mAs/frame
Displays R or Gy
Excel Add-in includes MQSA, Rad, and Fluoro templates
RS-232 computer interface
Enhanced dental capabilities
Introduction
The Victoreen® NERO® mAx, Non-invasive Evaluator of Radiation Outputs, evaluates the widest range of x-ray machines on the market today. Evaluation of pulsed fluoro, cine, computed tomography (CT), portable, mammographic, dental, radiographic, fluoroscopic, low, medium, and high frequency machines is finally made possible with a single instrument. This fifth generation instrument features 100 kHz sampling speed and direct mA/mAs measurements. The NERO mAx’s innovative Easy Flow Menu (EFM) system and flexible soft keys provide an intuitive, user-friendly operating environment for quick, accurate, and easy measurements.
All measurement modes and functions are displayed on the NERO mAx’s super bright LCD and are controlled by the five soft keys directly below the display and three hard keys to the right. Seven user selectable measurement modes and three system control modes are available and clearly displayed on the control console screen for easy access and selection.
Applications
The NERO mAx consists of the control console, detector, detector cable, two filter cards, mAs leads, Excel Add-in, AC
adapter, HVL plates, instruction manual, and carrying case.The compact control console houses the rechargeable battery, super bright easy to read backlit display, eight control buttons, and the sophisticated electronics necessary for accurate, reproducible measurements. Connectors for power input, RS-232, printer, scope output and the NERO mAx detector are located on the control console’s rear panel.
The NERO mAx detector contains sensors for simultaneously measuring kV, exposure or rate and mA or mAs. Solid state detectors are used to measure kV. An ion chamber, located in the top of the detector, is used for exposure/rate measurements. Connectors for external ion chambers and the NERO mAx detector interface are located on the rear panel
of the detector. The front panel has a keyed opening for the NERO mAx filter cards and a connector for mAs leads.The filter cards contain the various filters needed to accurately measure kilo voltage. Each filter card is coded so that the NERO mAx "knows" which filter is in use and its position. The NERO mAx also verifies that the filter card is valid for the selected measurement mode. The two filter cards are keyed so that they may only be inserted properly. The W/Al filter card and the Mammo filter card are clearly labeled as to the x-ray tube targets for which they are calibrated.
External chambers
External ion chambers for CT, mammographic, image intensifier tube, and special radiographic applications are available. Chamber calibration factors can be stored in the NERO mAx for direct readout of measurements. The Excel Add-in acquires measured data and waveforms directly into an Excel spreadsheet to maximize flexibility for report generation.
Product Specifications
Radio modeDescription Radio mode is used to make measurements on tungsten target, aluminum filtered radiographic x-ray generators Selections are available for % kV peak for the type of generator being tested. For example, Zero Crossing, Single Phase Pulse, 75%, 80%, or 90% of kVp modes are available for accurate exposure measurements on difficult x-ray machines
Simultaneously measures • kVp Avg
• exposure
• kV Eff
• exposure time
• kV Peak
• mAs
Mammo mode
Description Mammo mode is used to make measurements on mammographic x-ray generators
Simultaneously measures • kVp Avg
• exposure
• kV Eff
• exposure time
• kV Peak
• mAs
Flouro mode
Description Fluoro mode is used to make measurements on fluoroscopic x-ray generators. Fluoro mode supports both continuous fluoro and pulsed fluoro measurements
Continuous fluoro mode, measures • kVp Avg
• exposure rate (R/min)
• kV Eff
• mA
• kV PeakPulsed fluoro mode, measures
• kVp Avg
• exposure rate
(R/min and mR/pulse)
• kV Eff
• mAs/pulse
• kV Peak
AMSE mode
DescriptionAMSE mode is used for Automated Measurement of Sequential Exposures. This mode is used to measure the output of CINE generators
Measures• kVp Avg
• exposure rate
(mR/frame)
• kV Eff
• mAs/frame
• kV Peak
• time/frame (ms/frame)
CT Exposure mode
Description: CT Exposure mode is used to make CT exposure measurements using the Victoreen Model 6000-100 CT ion chamber. The CT probe must be connected to the NERO mAx detector’s external ion chamber input in this mode
Exposure mode
Description: Exposure mode is used to make exposure and rate measurements using the NERO mAx’s internal ion chamber or an external ion chamber
HVL mode
Description: In the HVL mode, the NERO mAx calculates half value layer based upon a series of exposure or rate measurements made with varying thicknesses of aluminum absorbers placed in the x-ray beam. A minimum of two exposures are required and up to ten exposures may be used
Calibrate mode
Description: Calibrate mode is used to enter and store calibration factors for ion chambers used with the NERO mAx
Setup mode
Description: Setup mode is used to setup various features of the NERO mAx. From the setup screen the user can set the instrument’s parameters such as the real time clock, temperature and pressure
Unit ID
DescriptionDisplays the NERO mAx’s serial number, firmware part number and level
Kilovoltage
DescriptionMeasured during the first 480 ms of exposure
Reproducibility
0.5 kV or ± 1%
RangeW/AI:30-60kV
W/AI:50-100KV
W/AI:80-160kV (1.2mm of AI)
Mo/Mo:22-35 kV (30 μ of Mo)Mo/Rh:22-40 kV(25 μ of Rh)Mo/AI:22-49kV(1mm of AI)
Rh/Rh: 25-49kV(25μ of Rh)Rh/AI:25-49kV(1mm of AI)
All caliberations are performed with NIST traceable caliberation beam.
Analyze/Display Cycle
Radio & mammo: 3 seconds for 0.1 second exposure, 1 second for each 32 ms of exposure timeFluoro & AMSE:
15 seconds for all exposures
Time
Radio mode: Measured during entire exposure at 90, 80, 75% rise/fall of waveform, zero crossing, or pulse countMammo mode:
Measured during entire exposure at 90% rise/fall of waveformAccuracy:
1 msResolution:
0.1 msRange:
All diagnostic exposures from 1 ms to 60 seconds
Exposure & Rate
Description: Measured during entire exposure with automatic energy, temperature, and pressure correctionMeasured quantity:
Roentgens or Grays, Accuracy: ± 5 %Reproducibility:
Radio and mammo modes: ± 2% or 2 mRResolution:
0.1 mRRange:
All diagnostic exposure and rate measurements from 1 mR to 9999 RFluoro rate:
0.1 R/min to 999 R/min
MAs and mA
Description: Measured invasively during entire exposureAccuracy: 2 %
Reproducibility: ± 1 % or 0.2 mAs
Range: 0.1 mAs to 999 mAs, 0 mA to 1000 mA
HVL
Accuracy: ± 5%Range: 0.1 to 99.9 mmAl
Physical
Display: Super bright 240 x 60 pixel, super twist LCD with cold cathode fluorescent backlightDetectors: Ion chamber and solid state (kV detectors)
Ion chamber volume: 45 cc nominal
Window area/density: 38 mg/cm2 Polycarbonate
HVL set: 2.30 mm, 1.0 mm, 0.3 mm Al
Power requirements: 12 VDC 1 A external supply. Rechargeable internal batteries supply more than 4 hours of continuous service with overnight charge
Size: Volume is 3960 in3 (0.065 m3)
Console: 9.00 x 9.12 x 3.25 in (22.86 x 23.17 x 8.26 cm)
Detector: 6.56 x 3.70 x 2.58 in (16.66 x 9.4 x 6.55 cm)
Filter cards: 2.4 x 6.25 x 0.31 in (6.1 x 15.88 x 0.8 cm)
Weight
Shipping: 23 lb (10.43 kg)Console: 4.56 lb (2.067 kg)
Detector (with card): 1.65 lb (0.747 kg)
Filter cards: 2.9 and 3.2 oz (0.090 and 0.094 kg
External chamber accessories
Radiographic Ion Chamber 30cm(Victoreen Model 6000-528): energy response: within 7% from 30 to 150 kVp (15 to 65 keV); cable: 15 ft (4.5 m); chamber dimensions: 4 x 4 x 0.54 in thick (10.2 x 10.2 x 1.4 cm)
Mammographic Ion Chamber, 3.3cm3
(Victoreen Model 6000-529): energy response: within 5% from 0.2 to 5.0 mm Al HVL (16 to 90 kVp); cable: 15 ft (4.5 m); chamber dimensions: 4 cm Ř x 1.5 cm thick. This chamber meets the needs of the MQSA for an external transparent chamber. Option: Probe Holder (Model 6000-529-95) for BRH2 test stand
Image Intensifier Ion chamber, 150 cm3
(Victoreen Model 6000-530B): energy response: ± 10% from 1.8 to 10 mm Al HVL; cable: 10 ft (3.0 m); chamber dimensions: 6.26 x 8 x 0.63 in (15.9 x 20.6 x 1.6 cm)
Scatter Ion Chamber, 400 cm3
(Victoreen Model 6000-532B): energy response: ± 5% from 32 to 662 keV; cable 10 ft (3.0 m); chamber dimensions: 6.26 x 8 x 0.63 in (15.9 x 20.6 x 1.6 cm)
CT Ion Chamber, 3.2 cm3
(Victoreen Models 6000-100 and 500-100): energy response: ± 5% from 1 to 10 mm Al HVL; cable: 3 ft (0.9 m); sensitive length: 4 in (10.0 cm); chamber inside Ř: 0.25 in (6.4 mm)
CT High Sensitivity Ion Chamber, 10cm3 , for multislice CT
(Victoreen Models 6000-200 and 500-200): energy response: ± 5% from 1 to 10 mm Al HVL; cable: 3 ft (0.9 m); sensitive length: 4 in (10.0 cm); chamber inside Ř: 0.45 in (11.44 mm)
Includes: control console detector detector cable two filter cards mAs leads Excel Add-in instruction manual HVL plates carrying case
AC adapter (specify from the following):
- USA, Japan: 110 VAC 12 VDC 1000 mA (part #14-238)
Europe: 230 VAC 12 VDC 1000 mA (part #14-401) UK: 230 VAC 12 VDC 1000 mA (part #14-414) Australia: 230 VAC 12 VDC 1000 mA (part #14-414 and 14-416)