Victoreen® ASM-990

 

Click Here to Secure Financing

 

 

 

   

ADVANCED SURVEY METER (ASM)

Victoreen® ASM-990

 

 

 

 

 


 

The ASM-990 is shown with
optional GM Pancake Probe
(Model 489-110D)
 

Probe is purchased separately
see below

Advanced survey meter for multiple hospital and environmental applications

Cell phone-like multi-function key for easy menu navigation

Backlit analog/digital display

Integrated handle

Error-free visual indication

Auto-ranging

Multiple probe use

Full-range audio output capability

Survey Mode data logging feature

Timed Peak Hold feature

Built-in scaler functions

Barcode scanner (optional)

Auto Power-Down feature extends battery life

 

Introduction

The ASM-990 is an easy-to-use instrument that detects alpha, beta, gamma, or x-ray radiation within an operating range of 1µR/h to R/h or 1 to 5,000,000 CPM, depending on the probe selection. Visual indication of measured values, as well as selectable parameters, are displayed on the analog/digital display.

The units are compatible with Geiger-Mueller (GM) detectors, proportional counters, and scintillation probes operating from 500 to 1300 volts.

Applications

The ASM-990 is designed to meet the high technology requirements of health physics, medical physics, and nondestructive testing applications. Radiation safety officers (RSO), nuclear medicine laboratories, diagnostic x-ray and hospital emergency room technicians, and environmental health physicists will appreciate this intelligent survey meter with its cell phone-like multi-function key for easy menu navigation.

The units, with purchased probe, are shipped calibrated and ready-to-use.

With the proper probe combination, this meter can be used as a general survey meter, and area monitor, a wipe est counter, and a contamination monitor.

The unit is supplied with a MHV connector to ensure compatibility with all Victoreen probes. A Model 992 is available that includes a fully calibrated internal energy compensated  1 R/hr GM detector.

The Model 993 features a fully calibrated internal pancake detector as well as an internal energy compensated 1 R/hr GM detector.

Features

  • Survey Mode feature allows user to store up to 5 separate sequences

  • Infrared Data (IrDA) port

LCD readout

  • 160 x 160 graphical LCD display shows digital average of the bar graph value

  • Analog scale has fifty-one elements arranged in a linear bar graph. Each element represents 2% of full scale. Scale markings are 0,2,4,6,8, and 10. Scale length is 2.2 inches (5.6cm)

  • Scale multiplier is 0.0001 to 1 million, depending on the probe selected and the units activated

  • LCD displays operational units such as radiation unit changes, response time changes, data, labels, and real time clock.

Pushbutton controls

  • Light: actives a background light for a preset or indefinite amount of time.

  • Start/Stop/Rst/Save: Saves current data; starts and stops recording of data; resets internal counter in scaler mode

  • Sel: activates the menu systems; allows user to choose particular settings, etc.

  • Up/Dn arrows: Enables user to navigate between various menus/selections

  • Ese: returns to previously selected menu or sub-menu

  • Audio: Allows the user to turn on/off the audio indicator.

 


 

Specifications
Operating modes

  • Rates

  • Integrate

  • Scaler (dual option): "Based On Measurement" or "Based On Time"

  • Timed Peak Hold

  • Data Logging

Operating ranges (dependent on selected probe)

µR/hr

mR/hr/hr

R/hr

µrem/hr

mrem/hr

rem/hr

µSv/hr

mSv/hr

Sv/hr

CPM

CPS

 

DPM99mTc

DPS131I

 

Bq125I

kBq123I

MBq201Tl

µCi67Ga

mCi18F

Ci57Co

 

µR

mR

R

µrem

mrem

rem

µSv

mSv

Sv

C(counts)

kC

MC

D(disintigrations)

kD99mTc

MC131I


Complementary units in the integrate mode with the integrated time value in seconds

Accuracy:
Within 10% of reading between 10% to 100% of full scale indication on any range, exclusive of typical energy dependence.
Accuracy is probe dependent.

Detector:
Accepts GM detectors and scintillation probes operating at high voltages between 500 and 1300 volts.

Environmental

Temperature range 14° to 122°F (- 10° to + 50°C)
Relative humidity 0 to 95%, non-condensing

Warm up time 5 second diagnostic check
Check source Natural uranium, mounted on the case
Power requirements Two “D” cells, 150 hours operation,
automatically indicates when battery is low
Housing material Proprietary polycarbonate, splash-proof case
Display Liquid crystal display, 2.2 x 2.2 in (5.6 x 5.6 cm)

Data logging modes:
The Log Data feature can easily be accessed via the Setup Sub-menu.

The ASM-990 can log/save a maximum of 500 data points in any of three separate modes
(Manual and Survey modes can utilize the optional barcode scanner):

Manual Individual data points can be saved by pressing the Start/Stop/Rst/Save button

Timed A data point will automatically be saved at user selectable time intervals in the range of 1 to 255 seconds.

Survey Programmed sequences can be accessed via the menu system. Pressing the Start/Stop Rst/Save button saves the current reading and displays the next survey location.

Programming of survey sequences, as well as retrieval of logged data, is accomplished via the built-in IrDA port. Label names up to 20 characters can be programmed into the ASM-990 to identify the individual survey locations.

Weight (without probe)

ASM 990, 992 2.1 lb (0.95 kg)
ASM 993 2.4 lb (1.09 kg)


Dimensions
4.125 (w) x 10.91 (d) x
2.5 in (h) (10.47 x 27.71 x 6.35 cm)


Probe connector The unit is
available with a MHV connector

The unit can be used with multiple probes
(5 total) by selecting the appropriate probe from the main menu. All calibration data for each probe is stored in the unit’s EEPROM

Specifications
for ASM-992 and ASM-993
internal energy compensated 1 R/hr GM
detector

Range 0.1 mR/hr to 1 R/hr

Radiation detected Gamma above 60 keV

Accuracy ± 10% of reading between 10% and 100% of full scale on
any range, exclusive of energy dependence

Typical energy dependence

Specifications for ASM-993 internal pancake detector

Radiation detected Alpha above 3.5 MeV, beta above 35 keV and
gamma above 6 keV

Range Background to 80 mR/hr

Window 15 cm2 (1.75 in Ø) mica, 1.4 to 2.0 mg/cm2

Typical background 30 CPM

Protective screen Stainless steel, hexagonal pattern providing 86%
open area

Accuracy ± 10% of reading between 10% and 100% of full scale on
any range, exclusive of energy dependence (protective cover open)

Typical energy dependence

Efficiency The internal pancake detector efficiency is shown below.
In a recent performance check, the numbers shown represent typical
results obtained:

Optional accessories
Mass Casualty Triage Tags, 500 per case (Model 990-ERT-101)

Barcoded Emergency Response Tags, 1000 per case
(Model 990-ERT-102)

Serial Port IrDA Adapter (Model 990-IR-SER)

Carrying Case (Model 990CC)

Wall Mounting Bracket (Model 990WM)

Universal Probe Holder (Model 990UPH)

Soft-Sided Holster (Model 990SH)

Shoulder Strap Assembly (Model 990SA)

Note: The shoulder strap assembly is only available for the ASM-993
and must be ordered with the instrument and factory installed.

Note: The ASM-990 series, with the customer selected probe is
calibrated to NIST standards. The ASM-990 series with GM probe
standard calibration is in R, Sv, and rems. Scintillation detectors are
calibrated in counts. Radionuclide specifi c effi ciency calibrations are
available upon request. For probe selection and calibration services,
see next page.


Available model(s)

For probe selection and calibration services, see next page.


 

Victoreen® Geiger-Mueller and Scintillation Probe Selection Guide

Pancake GM Probe
Model 489-110D

• Alpha above 3.5 MeV
• Beta above 35 keV
• Gamma and x-ray > 6 keV
• To 80 mR/hr
(800 μSv/hr)

Energy Compensated
GM Probe Model 90-12

• Beta above 200 keV
• Gamma and x-ray > 12 keV
• Up to 1 R/hr
(10 mSv/hr)

Thin End Window GM Probe Model 489-35
• Alpha above 4 MeV
• Beta above 70 keV
• Gamma and x-ray
> 6 keV
• Up to 80 mR/hr
(800 μSv/hr)

 

Utility 1 R/hr GM Probe Model 491-40
• Beta above 200 keV
• Gamma and x-ray
> 12 keV
• Up to 1 R/hr
(10 mSv/hr)

 

Gamma Scintillation Probe
Model 489-50
• Gamma and x-ray > 60 keV
• 1 x 1 in, 1.5 x 1.5 in and 2 x 2 in
NaI (Tl) detectors available

Alpha Scintillation Probe
Model 489-60
• Alpha above 4 MeV
• 1.5 in Ø ZnS (Ag)

Alpha/Beta Scintillation
Probe
Model 425-200

• Alpha above 350 keV
• Beta above 14 keV
• Plastic scintillator

Scintillation Pancake Probe
Model 489-200

• Beta above 100 keV
• Gamma and x-ray > 25 keV
• NaI (Tl) rectangular

Low Energy Gamma
Scintillation Probe
Model 425-110

• Gamma and x-ray > 10 keV
• NaI (Tl) 1 mm thick

100 cm2 Beta/Gamma
Scintillation Probe
Model 190-100BGS

• Beta 90Sr effi ciency 65.0%
• 126 cm2 active area
• Plastic scintillator

             

• Expert 24 x 7 same day emergency service
• Multi-unit contract pricing available
• Multi-vendor calibration available
• Radionuclide dependent calibrations

 

 

Calibration programs
ISO 17025:1999, ANSI Z540, Mammography MQSA, CNSC, NIST
& PTB Traceable

Quality programs
ISO 9001:2000, ISO 13485:1996, FDA/QSR, NRC/Part 50, Appendix B/Part

 


GM Pancake Probe
Victoreen
® Model 489-11OD

All purpose GM Pancake Probe alpha, beta, gamma and x-ray radiations

Used in nuclear medicine,
diagnostic x-ray, and geological and environmental surveys

Ideal for emergency response
teams


Features

High detection efficiency
 

Ergonomic design
 

Lightweight
 

Detachable probe cable
 

BNC or MHV connector
 

Easy to decontaminate


Introduction

The Victoreen Geiger-Mueller (GM) Pancake
Probe (Model 489- 110D) is a hand-held,
thin-window detector designed for alpha beta,
and gamma radiation measurements. It is
designed for use in conjunction with the Victoreen Models 990, 190, 290, and other standard GM
survey meters, and is configured for operating convenience in table top and floor surveys as well
as surveys of personnel and equipment. For storage and carrying ease, the probe fits into the standard handle clip on a survey meter.
 

To use the probe, attach the connector to the
probe input on the Survey meter and place
the window in proximity to the surface to be
surveyed. A screen protects the thin mica
window to facilitate rapid frisking of flat surfaces.
The open area of the hexagonal stainless steel
screen is 86%.
 

The GM probe comes in two configurations:
Model 489-llOD with an ABS plastic
housing, MHV connector, and foam grip, or
Model 489-11OE with a BNC connector.
This selection of connectors provides the ability
to attach the GM probe to most GM survey meters
on the market. Specify model number when
ordering. Replacement foam grip handles are available: Model 489-130-44. This same probe
design is available in a rugged metal housing as Model 489-11OC.

Applications
This all purpose GM Pancake Probe is used in a
wide range of applications where alpha, beta,
gamma and x-ray detection is required. Prime applications for this probe include nuclear medicine counter tops and frisker stations, leakage detection for low energy diagnostic x-ray machines, geological and environmental surveys or any place where there exists the suspicion that some form of radiation is present.
 

 