Supertech ®
Thousands of Products
Serving You is Our Pleasure
Product Categories
The Victoreen® X-Ray Test Device, Model 4000M+ does it all. Simply place the instrument in the x-ray beam, make one exposure, and it serially displays kVp Maximum, kVp Average, kVp Effective, dose, and time. The Model 4000M+ then automatically resets for the next exposure. A CsI photodiode pair provide the kVp measurements through five user-selectable filter pairs. This ensures optimum accuracy over the entire diagnostic range with minimum filtration dependence. Exposure measurements are made with a parallel plate ionization chamber located above the filter wheel. Exposure time is measured with quartz crystal accuracy. Plus, a variety of external ion chambers may be connected for even greater flexibility.
Key Features
- Measures kVp maximum, kVp average, kVp effective, dose and time in one exposure
- Compact, lightweight design
- Displays R or Gy
External ion chambers for Mammo, CT, image intensifier and photo timing measurements Automatic exposure reset for hands off operation Rechargeable Ni-Cd batteries provide more than six hours of continuous service RS-232 computer interface Storage scope output for real-time waveform display Reversible display for fluoro measurements
Product Specifications
Kilovoltage Accuracy: 1 kV Mo/Mo (22 to 35 kVp Mammo generators w/30 μ Mo Range: W/Al Tubes: 27 to 155 kVp Mo/Mo Tubes: 21 to 50 kVp
Time
Description: Measured during entire exposure at 90% rise/fall of
waveform
Accuracy: Within 2% or 2 ms, whichever is greater
Range: 1 ms to 10 seconds
Exposure
Description: Measure during entire exposure; kvp correctedAccuracy: ± 5%
Range: 10 mR to 10 R
Fluoroscopic
Rate
Description: Measured over one second intervals during fluoro exposureAccuracy: ± 5%
Range: 0.5 to 200 R/min
Detectors
kV: CsI/photodiode pair measures x-ray transmission through differential
attenuators
Time: Computed from kV waveform stored in memory against quartz
crystal time base
Exposure: Parallel plate ionization chamberVolume: 36 cm3
Window: 38 mg/cm2, 18.9 cm2 polycarbonate
Calibration: Reference to a NIST traceable voltage divider and a
calibrated exposure monitor during irradiation
Display
16 character dot-matrix LCD
Controls
Description: Model 4000+ has five rocker switchesOn/Off: Power switch
Radio/Fluoro: Select radiographic or fluoro operation
High/Low: Select for sensitivity
Roll: Roll thru data
Exposure/All: Select exposure only for external ion chamber
Mo/Mo or W/Al: Select anode/filter of x-ray tube
Connectors
Power: 9 VDC, 500 mAScope: BNC for oscilloscope connectio
RS-232: DB-9 connector configured as DCE. BNC and banana plug for
external Ion chamber
Power
Requirements
9 VDC 500 mA external supply. Rechargeable internal Ni-Cd batteries
supply more than six hours of continuous service with overnight charge
Dimensions
8.5 (w) x 9 (d) x 3 in (h) (21.5 x 23 x 7.6 cm)
Weight
Approximately 3.5 lb (1.59 kg)
HVL set
Aluminum filters: 2.3, 1.0, and 0.3 mm
External
Chamber
Accessories
Radiographic Ion Chamber, 30 cm3(Victoreen Model 6000-528): energy response: within 7% from 30 to 150
kVp (15 to 65 keV); cable: 15 ft (4.5 m); chamber dimensions: 4 x 4 x
0.54 in thick (10.2 x 10.2 x 1.4 cm)Mammographic Ion Chamber, 3.3 cm3
(Victoreen Model 6000-529): energy response: within 5% from 0.2 to 5.0
mm Al HVL (16 to 90 kVp); cable: 15 ft (4.5 m); chamber dimensions: 4
cm Ø x 1.5 cm thick. This chamber meets the needs of the MQSA for an
external transparent chamber. Option: Probe Holder (Model 6000-529-95)
for BRH2 test standImage Intensifier Ion Chamber, 150 cm3
(Victoreen Model 6000- 530B): energy response: ± 10% from
1.8 to 10mm Al HVL; cable: 10 ft (3.0 m); chamber dimensions:
6.26 x 8 x 0.63 in (15.9 x 20.6 x 1.6 cm)Scatter Ion Chamber, 400 cm3
(Victoreen Model 6000-532B): energy response: ± 5% from 32 to 662
keV; cable 10 ft (3.0 m); chamber dimensions: 6.26 x 8 x 0.63 in
(15.9 x 20.6 x 1.6 cm)CT Ion Chamber, 3.2 cm
(Victoreen Models 6000-100 and 500-100): energy response:
± 5% from 1 to 10 mm Al HVL; cable: 3 ft (0.9 m); sensitive length:
4 in (10.0 cm); chamber inside Ø: 0.25 in (6.4 mm)CT High Sensitivity Ion Chamber, 10 cm3, for multislice CT
(Victoreen Models 6000-200 and 500-200): energy response:
± 5%from 1 to 10 mm Al HVL; cable: 3 ft (0.9 m); sensitive length:
4 in (10.0 cm); chamber inside Ø: 0.45 in (11.44 mm)
Model Name
Power Adapter
4000M+
X-Ray Test Device with AC adaptor
(specify from the following)USA/Japan: 110 VAC 9 VDC 500 mA (14-301)
Europe: 230 VAC 9 VDC 500 mA (14-339)
UK: 230 VAC 9 VDC 500 mA (14-415)
Australia: 230 VAC 9 VDC 500 mA (14-415) and
adapter (14-416)
Related Products
4000EXL Toolkit for Excel RS232 Cable 190004 07-434 Ultra-High purity HVL Attenuators
Carrying Case
4000-69
The Victoreen® 4000 Toolkit for Excel is a complete software package for the 4000M+ NERO® that includes an Excel Add-In, called 4000 Add-In and Excel templates that may be used to evaluate the performance of radiographic, mammographic and fluoroscopic x-ray machines. The 4000 Add-In collects measured results from the 4000M+ NERO and places the data in the cells of the active Excel worksheet, starting at the active worksheet cell. The 4000 Add-In also may be used to acquire and graph radiation and kV waveforms from the 4000M+ NERO as well as remotely control the 4000M+ NERO.
Key Features
- Automatically collects measurement results and places them in an Excel worksheet
- Captures radiation and kV waveforms and charts them in an Excel worksheet
- Templates are provided for QA tests on radiographic, fluoroscopic, and mammographic X-ray machines
- Templates may be modified to perform user specific tests and generate user customized reports
- Allows complete remote control of the Model 4000M+ NERO • Complete on-line help speeds learning
- Compatible with Windows® 95, 98, ME®, NT® 4.0, 2000, and Microsoft® Excel 95, 97, 2000
- Automatically detects the presence of Model 4000M+ NERO
Product Specifications
Templates
Radiographic Template performs these tests:• Reproducibility• kVp accuracy• Timer accuracy• Linearity• Beam quality
Mammographic Template performs these tests:
• kVp accuracy• kVp reproducibility• Beam quality• Breast entrance exposure• AEC reproducibility• Average glandular dose• Radiation output rate
Fluoroscopic Template performs these tests:
• kVp accuracy• Beam quality• Fluoro exposure rate
Functions Controls
The 4000 menu provides an interface for the user to remotely control the 4000M+ NERO and retrieve radiation and kVwaveforms.
A description of each menu option follows:
Remote Control
Selects the 4000M+ NERO measurement mode and measurement options. This allows the user to select the correct measurement mode for the template in use. For instance, the radio mode is selected when using the radiographic template, the Mo/Mo target filter is selected when using the mammographic template and the fluoro mode is selected when using the fluoroscopic template
Retrieve Rad Waveform
Retrieves the radiation waveform data from the 4000M+ NERO and charts it in a new Excel chart
Retrieve kV Waveform
Retrieves the kV waveform data from the 4000M+ NERO and charts it in a new Excel chart
System requirements Microsoft Windows 95, 98, ME, NT 4.0, 200. Microsoft Excel 95, 97, 2000. One serial port (COM1 through COM4)
07-434 Ultra-High Purity HVL Attenuators
- HVL attenuators recommended for mammography
- 99.9% pure for accurate HVL
Because type-1100 aluminum is only 99.0% pure, it has some impurities that can give a HVL value which is 7.5% lower than those measured with pure aluminum. When doing HVL measurements with a mammography unit, it is recommended that the highest purity aluminum be used. This set of six attenuators satisfies this recommendation, because they are 99.9% pure (type-1145).
Specifications
Dimensions: 10 cm x 10 cm x 0.1 mm
Weight: 0.15 lb (0.06 kg)
Model Name
Description 07-434
Ultra-High Purity HVL Attenuators, set of 6
07-430
Standard Aluminum HVL Attenuators, set of 11
07-431
Copper HVL Attenuators, set of 10
Supertech ®
Address:
P.O. Box 186
Elkhart, IN 46515 USA
Sales: 800-654-1054
Phone: 574-264-4310
Fax: 574-264-9551
Email: sales@supertechx-ray.com