VictoreenM® 4000M+ X-ray Test Device
Non-Invasive

The Victoreen® X-Ray Test Device, Model 4000M+ does it all. Simply place the instrument in the x-ray beam, make one exposure, and it serially displays kVp Maximum, kVp Average, kVp Effective, dose, and time. The Model 4000M+ then automatically resets for the next exposure. A CsI photodiode pair provide the kVp measurements through five user-selectable filter pairs. This ensures optimum accuracy over the entire diagnostic range with minimum filtration dependence. Exposure measurements are made with a parallel plate ionization chamber located above the filter wheel. Exposure time is measured with quartz crystal accuracy. Plus, a variety of external ion chambers may be connected for even greater flexibility.

Key Features
Product Specifications
Kilovoltage Accuracy: 1 kV Mo/Mo (22 to 35 kVp Mammo generators w/30 µ Mo

Range: W/Al Tubes: 27 to 155 kVp Mo/Mo Tubes: 21 to 50 kVp
Time Description: Measured during entire exposure at 90% rise/fall of waveform

Accuracy: Within 2% or 2 ms, whichever is greater

Range: 1 ms to 10 seconds
Exposure Description: Measure during entire exposure; kvp corrected

Accuracy: ± 5%

Range: 10 mR to 10 R
Fluoroscopic Rate Description: Measured over one second intervals during fluoro exposure

Accuracy: ± 5%

Range: 0.5 to 200 R/min
Detectors kV: CsI/photodiode pair measures x-ray transmission through differential attenuators

Time: Computed from kV waveform stored in memory against quartz crystal time base

Exposure: Parallel plate ionization chamber

Volume: 36 cm3

Window: 38 mg/cm2, 18.9 cm2 polycarbonate

Calibration: Reference to a NIST traceable voltage divider and a calibrated exposure monitor during irradiation
Display 16 character dot-matrix LCD
Controls Description: Model 4000+ has five rocker switches

On/Off: Power switch

Radio/Fluoro: Select radiographic or fluoro operation

High/Low: Select for sensitivity

Roll: Roll thru data

Exposure/All: Select exposure only for external ion chamber

Mo/Mo or W/Al: Select anode/filter of x-ray tube
Connectors Power: 9 VDC, 500 mA

Scope: BNC for oscilloscope connection

RS-232: DB-9 connector configured as DCE. BNC and banana plug for external Ion chamber
Power Requirements 9 VDC 500 mA external supply. Rechargeable internal Ni-Cd batteries supply more than six hours of continuous service with overnight charge
Dimensions 8.5 (w) x 9 (d) x 3 in (h) (21.5 x 23 x 7.6 cm)
Weight Approximately 3.5 lb (1.59 kg)
HVL Set Aluminum filters: 2.3, 1.0, and 0.3 mm
External Chamber Accessories Radiographic Ion Chamber, 30 cm3

(Victoreen Model 6000-528): energy response: within 7% from 30 to 150 kVp (15 to 65 keV); cable: 15 ft (4.5 m); chamber dimensions: 4 x 4 x 0.54 in thick (10.2 x 10.2 x 1.4 cm)

Mammographic Ion Chamber, 3.3 cm3

(Victoreen Model 6000-529): energy response: within 5% from 0.2 to 5.0 mm Al HVL (16 to 90 kVp); cable: 15 ft (4.5 m); chamber dimensions: 4 cm Ø x 1.5 cm thick. This chamber meets the needs of the MQSA for an external transparent chamber. Option: Probe Holder (Model 6000-529-95) for BRH2 test stand

Image Intensifier Ion Chamber, 150 cm3

(Victoreen Model 6000- 530B): energy response: ± 10% from 1.8 to 10mm Al HVL; cable: 10 ft (3.0 m); chamber dimensions: 6.26 x 8 x 0.63 in (15.9 x 20.6 x 1.6 cm)

Scatter Ion Chamber, 400 cm3

(Victoreen Model 6000-532B): energy response: ± 5% from 32 to 662 keV; cable 10 ft (3.0 m); chamber dimensions: 6.26 x 8 x 0.63 in (15.9 x 20.6 x 1.6 cm)

CT Ion Chamber, 3.2 cm

(Victoreen Models 6000-100 and 500-100): energy response: ± 5% from 1 to 10 mm Al HVL; cable: 3 ft (0.9 m); sensitive length: 4 in (10.0 cm); chamber inside Ø: 0.25 in (6.4 mm)

CT High Sensitivity Ion Chamber, 10 cm3, for multislice CT

(Victoreen Models 6000-200 and 500-200): energy response: ± 5% from 1 to 10 mm Al HVL; cable: 3 ft (0.9 m); sensitive length: 4 in (10.0 cm); chamber inside Ø: 0.45 in (11.44 mm)


Model Name Power Adapter
4000M+ X-Ray Test Device with AC adaptor (specify from the following)

USA/Japan: 110 VAC 9 VDC 500 mA (14-301)

Europe: 230 VAC 9 VDC 500 mA (14-339)

UK: 230 VAC 9 VDC 500 mA (14-415)

Australia: 230 VAC 9 VDC 500 mA (14-415) and adapter (14-416)